1. Calibration
To calibrate the NutriSort device correctly, the operating temperature of the sorting station must be between 0°C and 40°C.
The weighing platform must be:
– mounted on a level and horizontal floor
– leveled
– pre-calibrated according to the factory calibration instructions
Press ENTER to enter the menu.
Click > until SERVICE MENU appears, then click ENTER.
Click > until TS4 appears. Press ENTER to start the calibration procedure
2. Scaling
Press ENTER to enter the menu.
Click > until SERVICE MENU appears, then click ENTER.
Click > until TS5 appears, then press ^v to change the value. Press ENTER to start the scaling procedure
After calibration is complete, click > until “Exit Menu” appears . Press ENTER to return to the main screen.